Abstract

We compared trysin-digested protein samples desalted by ZipTip(C18) reverse-phase microcolumns with on-plate washing of peptides deposited either on paraffin-coated plates (PCP), Teflon-based AnchorChip plates, or stainless steel plates, before analysis by matrix-assisted laser desorption/ionization-time of flight-mass spectrometry (MALDI-TOF-MS). Trypsinized bovine serum albumin and ovalbumin and 16 protein spots extracted from silver-stained two-dimensional gels of murine C(2)C(12) myoblasts or human leukocytes, prepared by the above two methods, were subjected to MALDI on PCP, AnchorChip plates, or uncoated stainless steel plates. Although most peptide mass peaks were identical regardless of the method of desalting and concentrating of protein samples, samples washed and concentrated by the PCP-based method had peptide peaks that were not seen in the samples prepared using the ZipTip(C18) columns. The mass spectra of peptides desalted and washed on uncoated stainless steel MALDI plates were consistently inferior due to loss of peptides. Some peptides of large molecular masses were apparently lost from samples desalted by ZipTip(C18) microcolumns, thus diminishing the quality of the fingerprint needed for protein identification. We demonstrate that the method of washing of protein samples on paraffin-coated plates provides an easy, reproducible, inexpensive, and high-throughput alternative to ZipTip(C18)-based purification of protein prior to MALDI-TOF-MS analysis.

DOI 10.1016/j.ab.2004.01.033